Controllers

WC77 Temperature Management Controller – Multiport

HotDog Patient Warming WC77 Temperature Management System - Multiport

Description

The Model WC77 – Multiport Controller provides maximum versatility and efficiency to patient temperature management, and brings the HotDog Patient Warming System to the forefront of advanced, patient-centered care. The aesthetics, touch screen interface, communication features, data feedback, and overall functionality are all designed to benefit the clinician and patient. It is compatible with older and current warming blankets, mattresses and cables.

Features:

  • Powers up to 4 blankets and a warming mattress simultaneously with individual temperature settings and safety alarms
  • Touchscreen user interface for easy operation
  • Integrated educational and training slide shows
  • Positioning guidance provided
  • Answers to FAQs and troubleshooting information accessible on-screen to simplify use
  • Adjustable settings for volume, brightness, temperature display, and language etc.
  • IV pole mount clamp with VESA standard interface for additional mounting optons
  • A clear cable retention loop/strap can be used to assist with cable management
  • Lightweight
  • Quiet operation
  • Three-year Limited Warranty
  • FDA 510K Cleared & CE Marked

Temperature Monitoring – Both models can display patient core temperature and output this temperature to a monitor (this feature is CE Marked and Pending FDA Clearance).

AutoMode – Temperature monitoring feature allows automatic control and adjustment of the warming devices based on the patient’s temperature (this feature is CE Marked and Pending FDA Clearance with temperature monitoring).

WC71 Temperature Management Controller – Single Port

Description:

The Model WC71 – Single Port Controller provides simplicity and efficiency to patient temperature management. It is ideal for the pre- and post-operative settings. The aesthetics, touch screen interface, communication features, data feedback, and overall functionality are all designed to benefit the clinician and patient. All older and current warming blankets, mattresses and cables are compatible.

Features:

  • Powers 1 blanket with individual temperature settings and safety alarms
  • Touchscreen user interface for easy operation
  • Integrated educational and training slide shows
  • Positioning guidance provided
  • Answers to FAQs and troubleshooting information accessible on-screen to simplify use
  • Adjustable settings for volume, brightness, temperature display, and language etc.
  • IV pole mount clamp with VESA standard interface for additional mounting optons
  • A clear cable retention loop/strap can be used to assist with cable management
  • Lightweight
  • Quiet operation
  • Three-year Limited Warranty
  • FDA 510K Cleared & CE Marked

Temperature Monitoring – Both models can display patient core temperature and output this temperature to a monitor (this feature is CE Marked and Pending FDA Clearance).

AutoMode – Temperature monitoring feature allows automatic control and adjustment of the warming devices based on the patient’s temperature (this feature is CE Marked and Pending FDA Clearance with temperature monitoring).

WC7X Specifications

Dimensions11” high x 7” deep x 8.75” wide (28 cm H x 17.8 cm D x 22.2 cm W)
Weight6.25 – 7.15 lbs (2.8 – 3.2 kg) excluding clamp
MountingCan be placed on  a horizontal flat surface (i.e. table top), clamped to an IV pole, or hung using a VESA mount of either FDMI MIS-C (35 × 75 mm) or FDMI MIS-D (75 × 75 mm) specifications
Temperature ControlMicro-processor
Blanket Operating TemperaturesBlanket Ports A  adjustable in 1oC increments
37° to 43° ± 1.0°C (98.6° to 109.4° ± 1.8°F)
Mattress Operating TemperatureMattress Port M adjustable in 1oC increments
35° to 40° ± 1.0°C  (95° to 104° ± 1.8°F)
Device RatingsInput: 100-240 VAC, 50/60 Hz, 850VA
Output A, B, C, D: 48 VDC, 480 VA Max each
Output M: 336 VA Max
Leakage CurrentMeets UL 60601-1 and IEC 60601-1 requirements for Class I, Type BF equipment.
Power Consumption850W maximum (during warm-up with all ports occupied)
CertificationsIEC 60601-1; EN 60601-1-2; UL 60601-1; CAN/CSA-C22.2, No. 601.1, EN 55011